Chatillon

Dinamometri e Strumenti per prove meccaniche

 

Chatillon® TCD110 - TCD225 Specification Comparison
Digital Force Testers

______________________________________________________________________________

 

 

 

The TCD110 shares the same TCD Console and the operation of the system is identical to the TCD225 System. The main differences are the frame’s capacity, speed and travel length. All accuracies are identical to the TCD225 System..

 

 

 

 

 

  TCD225 TCD110
Frame capacity 225 lbf
100 kg
1000 N
110 lbf
50 kg
500 N
Maximum Speed 50 in/min
1270 mm/min
20 in/min
510 mm/min
Minimum Speed 0.001 in/min
0.02 mm/min
0.001 in/min
0.02 mm/min
Speed Accuracy +1% Full Scale +1% Full Scale
Vertical Travel Distance 20 inch
510 mm
Weight (with Console & Sensor)
Weight (with Console & Sensor) 72 lbs
33 kg
65 lbs
30 kg
Shipping Weight (approx.) 90 lbs
41 kg
82 lbs
37 kg
Load Measurement System - Accuracy better than 0.1% Full Scale
- Calibrate on-site to better than 0.25% of Reading
- Measuring System meets or exceeds the following: ASTM E4, BS 1610, DIN 51221, ISO 7500-1
Strain Measurement System - Position accuracy better than 0.001 inch (0.02mm) or 0.05% displacement (whichever is greater)
- Strain measurement meets or exceeds the following: ASTM E83, BS 3846, ISO 9513
- Deflection compensation is standard
Data Sampling - Up to 1000 Hz
- Selectable Filtering from 1-1000 Hz
Data Outputs - USB Serial Data
- RS232 Serial Data
- Save data to USB flash drive or interface to USB serial devices
Conformance - BS EN 61010 Safety Requirements for Electrical Equipment
- BS EN 61000 EMC Generic Emissions Standard
- BS EN 61000-6 EMC Generic Immunity Standard
Environmental - Operating Temperature (40o to 110oF, 5o to 45oC)
- Storage Temperature (0o to 130oF, -17o to 54oC)
- Relative Humidity 20 to 85% Non-condensing

 

 

 

For more info visit the product pages at:

 

 

 

 